| Tender ID | 3781705 |
|---|---|
| Organisation Name | Medical Health And Family Welfare Department |
| Tender Brief | Supply, Installation, Testing and Commissioning Of Medical Gas Pipe Line System And Modular Operation Theater Work In District Health Facilities Comes Under Medical And Health Division Kota Sub District Hospital Anta, Atru And Mangrol |
| Last Date For Submission | 1st Apr, 2026 |
| Opening Date | 02 Apr, 2026 |
| Document Cost | 5000.00 |
|---|---|
| EMD | 926000.00 |
| Estimated Cost | 46300000.00 |
| City | Kota |
|---|---|
| State | Rajasthan |
| # | Item | Qty | Units |
|---|---|---|---|
| 1 | Wall paneling system "SS Sandwich : It should be 0.8mm 304 Grade Stainless Steel sandwich (both side 0.8mm sheet) panel with core consisting of rigid polyurethane foam, which has been injected under high pressure, with a minimum density of 40 kg/m3. Cladding structure insulated Stainless steel wall panels with Total Panel thickness 50-60mm. "The individual wall panels shall use the tongue and groove technology/Suitable technology for joining two panels, no welding should be allowed. The gaps between panels shall be suitably filled with metal filler/epoxy and sanded flush or medical grade silicone/ Monolithic sealing in case of SMS "Stainless Steel plate finished to fine grain surface, treated properly to take antifungal paint (Not Applicable in case of SMS Option)" Paneling should be easy to maintain, durable, antistatic/conductive and fire retardant. Clearance between inner panel and outer wall should be sufficient to allow the maintenance personnel for service. This closed space should be flushed continuously to eliminate dust and bacterial accumulation. Anti bacterial paint should be coated on the wall (Not Applicable in case of SMS Option) Bidder should maintain anti-bacterial paint during warranty and CMC period.(Not Applicable in case of SMS Option) Wall elements should be resistant to all standard cleaning agents, disinfectants and fumigation agents. Panel should be covered with protective sheath to prevent scratch during installation. It should have minimum number of junction. The junction should be seamless and should be sealed with suitable sealants. The wall panels should be CE/UL Listed/BIS/DIN 1.4301 certified (Not Applicable in case of SMS Option) Third party test certificate for SS/ SMS from material testing lab. (Govt. authorized) - to be provided at the time of pre dispatch inspection/supply. | 400.00 | sqm |
| 2 | Ceiling Paneling System The prefabricated Operating Room Should be of SS Sandwich. It shoud be 0.8 mm 304 grade stainless steel sandwich (both side 0.8 mm sheet) panel with core consisting of rigid polyurethane foam, which has been injected under high pressure, with a minimum density of 40 kg/m3. Cladding structure insulated stainless steel wall panels with Total panel thickness 30-40 mm. Support elements: Suspension bracket with tension spring/ threaded rod Material: High quality galvanized or powder coated steel. Room lighting, air supply inlet, ceiling service units, return air outlets, etc should be integrated with SS metal ceiling system. The individual panels except those at the edges should be removable individually. The ceiling material should be CE/ UL/BIS /DIN 1.4301 certified (Not Applicable in case of SMS Option) Anti bacterial paint should be coated on the ceiling.(Not Applicable in case of SMS Option) Third party test certificate for SS/ SMS from material testing lab. (Govt. authorized) - to be provided at the time of pre dispatch inspection/supply. | 160.00 | sqm |
| 3 | PVC Flooring with self leveling It should be with 2mm antistatic seamless PVC flooring Floor should be smooth, non-slip, impervious material conductive enough to dissipate static electricity but not conductive enough to endanger personnel from electric shock. Electrostatic charge dissipation combat PVC seamless flooring of very high quality should be provided. Thickness not less than 2 mm. Continuous roll should be used and joints should be welded by special PVC thermal welding units using PVC welding bars of same colour The sheets should be highly durable with resistance to shock and indentation. It should be scratchproof also. The conductive material should be uniformly impregnated as grains. It should be inert to body fluids, chemicals and disinfectants. Should not be affected by temperature variation within the OT. The floor should efficiently discharge electric charges up to 2 kV Flooring should be done by skilled workers of accredited agencies authorized by the supplier of PVC sheets. The electrical resistance (point to ground) should be within 2.5x10^4 to 5x10^6 ohms. The floor should not allow build up of electrical charge beyond 100 volts due to antistatic effect. The corners should not be terminated sharply and concealed cove- former (aluminium) should be used to overlap the wall panel to a height of approx.25mm and sealed perfectly and uniformly. Self leveling compounds should be used. The conductive copper grid laid underneath the PVC sheet should be supported by liquid epoxy compounds allowed to set as a uniform and level surface. The copper strips to be made visible by grinding and no copper strip should project more than 0.5mm above level surface to avoid damage to the PVC sheet. One earthing lead should be brought out from every 150sq.ft area and attaching it to the main earthing strip/ground. Copper grounding strips (0.05 mm thick, 50 mm width) should be laid flat on the floor in the conductive adhesive and connected to copper strip of grounding. The connection from copper grid should be brought out uniformly at places to form equipotential grid. Flooring should be mechanically shock proof, scratch proof, flame retardant and anti microbial Corners should be uniformly curved Final surface should be non corrosive to biological fluids and detergents. Colour should be uniform pleasant and matching with ambience and as approved by respective consignee. Suitable self-leveling should be done before PVC flooring to avoid undulation within the MOT. | 200.00 | sqm |
| 4 | Sterile Anti-bacterial Coating - (cladding type OT) "Anti-bacterial and Anti-fungus sterile coating will be applied on the SS finished walls & ceiling with at least 2 3 coats to achieve a thickness of minimum 200 micron. Before application of sterile wall coating, an appropriate primer will be applied. The primer will be supplied along with the paint and will be imported. The primer will be applied prior to coating of Anti-bacterial & Anti-fungus sterile paint on SS walls & ceiling. 2 to 3 coats will be applied to achieve a thickness of 200 microns. Sterile coating and primer will be applied on the surface by air-less spray gun. Sterile coating will be able to withstand commonly used cleaners in the operation rooms and even withstand regular steam cleaning." | 560.00 | sqm |
| 5 | HERMETICALLY SEALED DOOR Door sizes should be as per below option and quantities will be as per BOQ of respective institute HERMETICALLY SEALED DOORS Size 2.1mx1.8m HERMETICALLY SEALED DOORS Size 2.1mx1.8m with Lead Line(As per AERB Norms) This should be a hermetically sealed, single sliding door of 2.1 (H)X 1.8 m(W) The controller should be capable of being operated by elbow switches/foot switches as well as touch less sensor. The track should be of stainless steel/Aluminum and the running surface for the top rollers should be suitably angled to reduce resistance to movement The door leaf should be hung by means of hard plastic rollers of high quality with double bearing at the top. Rollers should be provided under the stainless steel/Aluminium track to enable smooth and noiseless movement. Opening and closing of the door should be microprocessor controlled electromechanical movement. The door material should be of SMS/ HPL Color should match the interior and care should be taken to make the leaf strong and light weight. One should be able to open and close the door effortlessly incase of failure of automatic mechanism. Door opening handle should be strong and sturdy. Material should be of SS (gloss/matte finish). Should be provided with high quality cylindrical/ ESPg lock. Door leaf should have high quality synthetic rubber gasket with long life to ensure hermetic sealing (to maintain air pressure differential). Air tightness 99.99% at a pressure of minimum 75Pa The finished floor on either side of the door should be perfectly level (maximum permissible difference +1mm). The overall thickness of the finished door should be 30-60mm. The inner part of the door should be filled with CFC free polyurethane foam thickness of 48mm or nearby. (Sealed airtight to prevent further ingress of any microbial organism). The door and controls should comply with IEE regulation. All motors used should be DC brushless/PMDC motors with essential isolation from mains. Door should be with vision window 300 mm x 300 mm with double glazed panels and hermetically sealed. Door movement should have minimum noise. The starting time after receiving the signal should be adjustable between 0.5 to 20 seconds. The door controller should be CE marked. Test certificate for hermetically sealed door frame (factory test certificate) should be enclosed with the pre dispatch documents. Window with Motorized Blinds for MOT : size approx. 1.5m x 1m and 2m x 1.5m | 4.00 | Each |